English
Language : 

10EK3M Datasheet, PDF (1/2 Pages) Delta Electronics, Inc. – POWER ENTRY MODULE FILTERS
EKSERIES POWER ENTRY MODULE FILTERS
INTRODUCTIONS
SPECIFICATIONS
1. EK Series filters are for UL 544 & UL 2601 health care equipment.
2. All part numbers are UL recognized, CSA certified
and VDE approved.
COMPONENTS
PART NO.
03EK3M/AM
03EK3SM/AM
03EK4M/AM
06EK3M/AM
06EK3SM/AM
06EK4M/AM
10EK3M/AM
10EK3SM/AM
10EK4M/AM
L (mH)
8
2.5
1.6
C1 (uF)
0.22
Lg: 400uH (EK3AM, EK3SAM only)
C2 (uF)
0.1
R (MW)
1
1. Maximum leakage current each
line-to-ground: @ 115VAC 60Hz: 2.0uA
@ 250VAC 50Hz: 5.0uA
2. Hipot rating (one minute)
line-to-ground: 2250VDC
line-to-line: 1450VDC
3. Operating frequency: 50/60Hz
4. Rated voltage: 115/250VAC
ELECTRICAL SCHEMATIC
SW
L
L
I
N
E
N
G
C1 R
L
FIG. A
F
C2
L
O
A
D
G'
1
2
110V
3
110V
4
MINIMUM INSERTION LOSS IN dB
COMMON MODE (L-G) IN 50 OHM SYSTEM
CURRENT
RATING
FREQUENCY-MHz
0.05 0.10 0.15 0.50 1.0 5.0 10 30
3A (M)
22 27 32 42 36 20 15 6
6A (M)
* 10A (M)
3A (AM) (SAM)
12 18 20 25 25 20 13 6
8 10 14 20 25 20 17 10
22 27 32 30 36 22 18 6
6A (AM) (SAM) 12 18 20 25 25 20 13 6
*10A (AM) (SAM) 8 14 24 20 25 20 17 12
DIFFERENTIAL MODE (L-L) IN 50 OHM SYSTEM
3A (M)
2 4 18 48 55 50 50 35
6A (M)
4 4 12 38 50 50 50 25
*10A (M)
4 6 6 30 45 50 50 35
3A (AM ) (SAM) 35 35 45 50 55 40 25 10
6A (AM) (SAM) 25 45 25 40 50 40 25 10
*10A (AM) (SAM) 20 28 38 45 50 40 25 13
*CSA approved 6A at 125/250VAC. VDE approved 6A at 250VAC.
SW
L
L
I
N
E
N
G
Lg
C1 R
L
FIG. B
L
L
I
N
C1 R
L
E
N
G
FIG. C
L
L
I
N
E
N
G
Lg
C1 R
L
FIG. D
FOR XXEK3SM
1
F
2
C2
L
O
A
D
3
G'
4
FOR XXEK3SAM
1
F
2
C2
L
O
A
D
3
G'
4
FOR XXEK4SM
1
F
2
C2
L
O
A
D
3
G'
4
FOR XXEK4SAM
110V
110V
110V
110V
110V
110V
DELTA PART NO.
* FUSE
IEC CONNECTOR
**DOUBLE-POLE POWER SWITCH
VOLTAGE SELECTOR SW
ELECTRICAL SCHEMATIC
03EK3M
06EK3M
10EK3M
DOUBLE
FIG. A
03EK3AM
06EK3AM
10EK3AM
DOUBLE
03EK3SM
06EK3SM
10EK3SM
SINGLE
03EK3SAM
06EK3SAM
10EK3SAM
SINGLE
FIG. B
FIG. A
FIG. B
03EK4M
06EK4M
10EK4M
DOUBLE
_
FIG. C
* UL, CSA & VDE approved
Current rating: UL-10A both at 125VAC & 250VAC.
CSA & VDE-6A at 250VAC.
* * UL, CSA & VDE approved
Current rating: UL & CSA-10A both at 125VAC & 250VAC.
VDE-10A at 250VAC.
Electrical lifetime: 10,000 cycles.
Maximum inrush current: 51A.
7-10
03EK4AM
06EK4AM
10EK4AM
DOUBLE
_
FIG. D
03EK4SM
06EK4SM
10EK4SM
SINGLE
03EK4SAM
06EK4SAM
10EK4SAM
SINGLE
_
_
FIG. C
FIG. D