English
Language : 

F3099AA06 Datasheet, PDF (1/1 Pages) List of Unclassifed Manufacturers – Designed to Meet UL544 and IEC601 Specifications for Medical and Dental Equipment
F3099 RFI Filters
Features:
• Designed to Meet UL544 and IEC601 Specifications
for Medical and Dental Equipment, both patient care
and onpatient categories.
• Leakage current in this series is extremely low to
satisfy the stringest leakage current limit imposed by
safety regulations for medical and dental equipment.
Specifications:
Maximum Voltage: 250VAC Maximum - 50/60 Hz
Rated Current: 250VAC
6A
Current Overload: 6X for 8 seconds
Hi-Pot Test (1 min):
Line to Ground 1500VAC
Line to Line
1450VDC
Insulation Resistance: 9 x 109 Ω at 100VDC
Ambient Temperature: 40°C Max at rated current
Humidity Range: 0% to 95% R.H.
Termination:
A: Quick Connect
B: Wire
Maximum Leakage Current:
Each Line to Ground F3000 Series
115VAC, 60Hz:
2 μA
250VAC, 50Hz:
5 μA
Agency Approvals:
F3099AA (6Amp) Dimensions
Amps
6A
A
2.53
(64,30)
2.53
(64,30)
B
1.82
(46,2)
1.82
(46,2)
C
0.78
(19,8)
0.78
(19,8)
D
2.126
(54,0)
2.126
(54,0)
E
2.53
(64,30)
1.32
(33,5)
6.36(0.25) FASTON TERMINAL
WITH Ø1.78 (0.070) HOLE
(SLOTTED HOLE
FOR GROUND)
AD
B
F3099 Series
Simplified Schematic
L
L
G
N
N
BLUE
AD
BROWN
WIRE LEADS, 18AWG,
(16AWG FOR 10A).
102(4.0) MIN.
BLUE
G/Y
B
BROWN
E
C
E
C
Nominal
Current
Rating
Part
Number
Termination
Line/Load
MODE
MINIMUM INSERTION LOSS - dB (50 ohm Circuit)
Frequency - MHz
0.05 0.10 .15 .50 1.0 5.0
10
30
6A
F3099AA06
F3099BB06
QC/QC
Common
3
7
11
20
22
24
22
18
ø
Differential
3
6
14
20
30
35
35
35
NOTE: Other combinations of terminals may be specified on special order.
72 www.curtisind.com
Curtis Industries
A Division of Powers Holdings, Inc.
Specifications subject to change.
Dimensions are shown for
reference purposes only.