CDB5467U Datasheet, PDF (18/38 Pages) Cirrus Logic – CDB5467U Engineering Board and GUI Software
CDB5467U
2.7 Pulse Rate Window
The CS5467 features a pulse-rate energy output. The CDB5467U has the capability to demonstrate the
functionality of this output in the Pulse Rate Output Window. See Figure 11. The Pulse Rate Output Win-
dow can be accessed by pressing <F5>, or by pulling down the Menu menu, and selecting the Pulse Rate
Window item.
Figure 11. Pulse Rate Output Window
2.7.1 Integration Seconds
This field allows the user to select the length of time over which pulses will be collected after the Start
button is pressed. Because there is a time delay (2x N / OWR) from start conversion to pulse output, the
Integration Seconds field is normally set to a delay longer than the actual pulsing period.
2.7.2 Periods To Average
This field allows the user to average a number of integration periods.
2.7.3 Start Button
When the Start button is pressed, the CDB5467U will capture pulse rate data according to the values in
the Integration Seconds and Periods to Average fields. After each integration period, the Pulse Count,
Frequency, Average Freq., and Standard Deviation columns will be updated. The Average Frequency and
Standard Deviation columns will only be updated after all the integrations have been collected. The soft-
ware stops collecting data when the user presses the Stop button, or when the data collection is finished.
Due to speed limitations of the on-board microcontroller, some higher pulse rates cannot be accurately
collected.