English
Language : 

AHP6301 Datasheet, PDF (1/6 Pages) Astrodyne Corporation – 63W Medical Power Adapters for Health Care Applications
63W Medical Power Adapters for Health Care Applications
AHP6301 series
Wide 80-275VAC Input
medical Medical Safety IEC60601-1 3rd Edition; 2MOPP
Choice of IEC320-C14, C8 or C6 Input Receptacles
Single Outputs froHm 1P2 Uto 468 V3DBC series
63
EfficieThnecyHPLUe6v3eBlsVeries of AC/DC switching mode power supplies
provide 63 Watts of continuous output power . All supplies are
ClassUCIILS9oP4RrV-C1-11lamEsNins5c5IoI0m11plcilaansts.
All models
B emission
meet FCC Part-18 class B and
Limits and are designed to
comply with ANSI/AAMI ES 60601-1: 2005(UL/cUL 3rd Edition), EN
60601-1:2006 (TUV/T-mark 3rd Edition) and new CE requirements.
All units are 100% burned in and tested.
RoHS2
2011/65/EU
FEATU
* Wide O
* IEC-32
* Single
* Over V
* Medic
* Input t
* Energy
* Class II
* 3 year
Model Number Output Watts Output Voltage Output Amps Efficiency Total Regulation
*AHP6301(A,B,C)-05 63
*AHP6301(A,B,C)-06 63
*AHP6301(A,B,C)-07 63
*AHP6301(A,B,C)-08 63
*AHP6301(A,B,C)-09 63
*AHP6301(A,B,C)-10 63
*AHP6301(A,B,C)-11 63
12VDC
5.25
15VDC
4.20
18VDC
3.50
24VDC
2.62
30VDC
2.10
36VDC
1.75
48VDC
1.31
84%
±5%
85%
±5%
85%
±5%
85%
±3%
86%APPROV±A3%LS:
86%
±3% CB
86%
±3%
EN 60601-1
PS
IEC 60601-1
E
* Note:
Electrical Characteristics:
Symbol
Characteristic
APPLIC
* Home
* Patien
* Blood
* Portab
* ECG m
GENER
* Short C
* Coolin
* Flamm
* Protec
* Safety
(TUV/T-m
The Asterisk * denotes a choice of receiving this product with an IEC-320-C14 3 Pole Input RecVeipntascle (SAa)foertywAitphparnovIEalCIn-3p2u0t-VCo8lt2agPeoRleaInngpeut RSeacfeepttyaAclpep(rBo)voarl &wiSthpecification i
an IEC-320-C6 3 Pole Input Receptacle (C). See the Mechanical files at the end of this specificVatiinon for cInlapruityt.OSpimerpaltyeaVdodlttahgeeleRtatenrg“eA”, “B” or “CD”eatfateilrto“AsHePe6F3ig0.1 i(nDerate linearly
the model number (i.e. “AHP6301A-05” for C14 or “AHP6301B-05” for C8 or “AHP6301C-05” foFriC6) wheInnpourtdFerreinqguetonrceyceive the desired versioSnin. e wave
Po
Output Power Range
See Rating Chart
Iil
Low Line Input Current
Full Load, Vin=100VAC
Iih High Line Input Current
Full Load, Vin=240VAC
Irl
Low Line Input Inrush Current
Full Load, 25°C, Cool start, Vin=1
Irh High Line Input Inrush Current
Full Load, 25°C, Cool start, Vin=2
Efficiency
Full Load, Vin=230VAC, Detail to
△Voi Line Regulation
Full Load, Vin=100~120VAC or 20
OVP Over Voltage Protection
OLP Over Load Protection
Recovers automatically after fau
ttr Time of Transient Response
Full Load, Vin=110VAC
thu Hold-Up Time
Full Load, Vin=100VAC
ts
Start-up time
Full Load, Vin=100~240VAC
Tc
Temperature Coefficient
All Condition
HV Dielectric Withstanding Voltage (P-S) Primary to Secondary, limit curre
WWW.ASTRODYNETDI.COM
EMI EMC Emission ASTRODYNE USA: 1-C8o0m0p-8li2an3c-8e0to82EN55011 (CISPR1
ASTRODYNE PACIFIC: 886-2-26983458
Environmental:
Symbol
Characteristic