English
Language : 

LPS50-M Datasheet, PDF (1/2 Pages) Astec America, Inc – 60 Watts
60 Watts
LPS50-M Series
Total Power:
Input Voltage:
# of Outputs:
55-60 Watts
90-264 VAC
127-300 VDC
Single
Special Features
• Medical Safety Approvals
• Universal input
• 2” x 4” footprint
• Remote sense
• Overload and short circuit protection
• Adjustable Output Voltage
• High efficiency
• High MTBF
• Built in EMI filter (CISPR 22 Class B)
• LED power good indicator
• Input power <74 watts
• Complies with EN61000-3-2
• UL Class I approved
• UL Class II available, consult factory
Electrical Specs
Input
Input range
Frequency
Inrush current
Input power
Efficiency
EMI RFI
Safety ground
leakage current
Output
Maximum power
Adjustment range
Hold-up time
Overload protection
Overvoltage protection
Remote sense
90-264 VAC (wide range)
127-300 VDC
47-440 Hz
<60 A peak @ 230 VAC,
cold start @ 25°C
<74 Watts
80-85% typical at full load
FCC Class B conducted
CISPR 22 Class B conducted
EN55022 class B conducted
VDE 0878 PT3 class B conducted
275 uA @ 50/60 Hz, 264 VAC input
60 W for convection (LPS52, 55W)
±20% minimum
10/20 ms 115/230 VAC Input line
Short circuit protection on all outputs.
Case overload protected @ 110-160% of
normal rating
30-50% above nominal output
Compensates for 0.5 V lead drop max.
Will operate without remote sense
connected. Reverse connection
protected.
Environmental
Operating temperature: 0° to 50°C ambient derate
each output as 2.5% per degree from 50° to 70°C.
-20°C start up.
Storage temperature: -40°C to +85°C
Electromagnetic susceptibility:
designed to meetEN61000-4;
-2, -3, -4, -5, -6, -8, -11 Level 3
Humidity: Operating; non-condensing 10% to
95%RH
Vibration: IEC68-2-6 to the levels of IEC721-3-2
MTBF demonstrated: >550,000 hours at full load
and 25°C ambient conditions.
Safety
UL
UL 60601-1
CSA CSA-C22.2 No. 601.1-M90
VDE EN60601-1
CE
LVD & EMC
74
Americas: 1-888-41-ASTEC
Europe (UK): 44 (1384) 842-211 Asia (HK): 852-2437-9662
rev 08.16.05