English
Language : 

LPS40-M Datasheet, PDF (1/3 Pages) Astec America, Inc – 40 Watts Medical
Embedded Power for
Business-Critical Continuity
LPS40-M Series
40 Watts
Medical
Total Power:
Input Voltage:
# of Outputs:
40-55 Watts
85-264 VAC
120-300 VDC
Triple
RoHS
Rev. 04.14.06
LPS40-M Series
1 of 3
Special Features
• Medical safety approvals
• Universal input
• 3" x 5" footprint
• Less than 1U high
• Remote sense on main output
• Low output ripple
• Adjustable 5V output
• Overvoltage protection
• Overload protection
• 110 KHz switching frequency
• LPX40 enclosure kit available
Safety
VDE 0750/EN60601-1 (IEC601)
UL
UL2601 E182560
CSA CSA 22.2 No.601.1
LR109492
CE Mark (LVD)
Electrical Specifications
Input
Input range
85-264 VAC; 120-300 VDC
Frequency
47-440 Hz
Inrush current
<18A peak @ 115VAC; <36A peak @ 230VAC, cold start @ 25°C
Input current
1A max. (RMS) @ 115VAC
Efficiency
70% typical at full load
EMI filter
FCC Class B conducted; CISPR 22 Class A conducted;
EN55022 Class A conducted; VDE 0878 PT3 Class A conducted
Safety ground
leakage current
Output
<75μA @ 50/60 Hz, 264 VAC input
Maximum power
40W for convection; 55W with 30CFM forced air
Adjustment range
-5, +10% minimum
Hold-up time
20ms @ 40W load, 115VAC nominal line
Overload protection Short circuit protection on all outputs
Case overload protected @ 110 -145% above peak rating
Overvoltage protection 5V output; 5.7 to 6.7VDC. Other outputs 10% to 25%
above nominal output
Remote sense
Compensates for 0.5 V lead drop minimum.
Will operate without remote sense connected.
Reverse connection protected.