LPS170-M Datasheet, PDF (1/2 Pages) Astec America, Inc – 175 Watts
175 Watts
LPS170-M Series
MEDICAL
Special Features
• Medical safety approvals
• Active power factor correction
• IEC EN61000-3-2 compliance
• Wide Range Adjustable output
Remote sense on main output
• Single wire current sharing
• Power fail and remote inhibit
• Built-in EMI filter
• Low output ripple
• Overvoltage protection
• Overload protection
• Thermal overload protection
• DC power good
• 5 V standby output
• 12 V Aux output
• Optional cover (-C suffix)
Environmental
Operating temperature: 0° to 50°C ambientderate
each output at 2.5% per degree from 50° to 70°C
Electromagnetic susceptibility: designed to meet
IEC EN61000-4, -2, -3, -4, -5, -6, -8, -11, Level 3
Humidity: Operating; non-condensing
5% to 95%
Vibration: Three orthogonal axes, sweep at
1 oct/min, 5 min. dwell at four major resonances
0.75 G peak 5 Hz to 500 Hz, operational
Storage temperature: -40° to 85°C
Temperature coefficient: ± .04% per °C
MTBF demonstrated: >550,000 hours at full load
and 25°C ambient conditions
Total Power: 110-175 Watts
Input Voltage: 85-264 VAC
120-300 VDC
# of Outputs: Single
Electrical Specs
Input
Input range
Frequency
Inrush current
Efficiency
EMI filter
Power factor
Safety ground
leakage current
85-264 VAC, 120-300 VDC
47-63 Hz
38 A max, cold start @ 25°C
75% typical at full load
Meets FCC Class A conducted
CISPR 22 Class A conducted
EN55022 Class A conducted
VDE 0878 PT3 Class A conducted
0.99 typical
<250 µA @ 50/60 Hz, 264 VAC inputS
Output
Maximum power
Adjustment range
Hold-up time
Overload protection
Overvoltage protection
Standby output
Aux output
110 W convection (75 W with cover)
175 W with 30 CFM forced air
(130 W with cover)
2:1 wide ratio minimum
20 ms @175 W load at nominal line
Short circuit protection on all outputs.
Case overload protected @ 110-145%
above peak rating
10% to 40% above nominal output
5 V @ 2A regulated ±5%
12 V @ 1 A -5 %, +10%
Logic Control
Power failure
Remote inhibit
Remote sense
DC Ok
TTL logic signal goes high 100 - 500
msec after V1 output; It goes low at
least 4 msec before loss of regulation
Requires contact closure to inhibit
outputs
Compensates for 0.5 V lead drop min.
Will operate without remote sense
connected. Reverse connection
protected.
TTL logic signal goes high after main
output is in regulation. It goes low
when there is a loss of regulation
Safety
VDE
0750/EN60601-1 (IEC601)
UL
UL2601
CSA
CSA 22.2 No. 601.1
CE
Mark (LVD)
100 America’s (USA): 1-888-41-ASTEC
Europe (UK): 44 (1384) 842-211
Asia (HK): 852-2437-9662
rev 06.23.04