3239 Datasheet, PDF (3/8 Pages) ASM GmbH – DIGITAL HiTESTER
2
Minimizing tact time
with sequence control
High-speed comparator and external input/output
■ EXT.I/O
Loading of saved settings from panel……LOAD 0 to 4
Measurement start trigger input…………TRIG
Measurement end signal output…………EOC
Comparator output…………………………Hi, IN, Lo
Internal power supply +5 V (max. 50 mA)……INT. DC V
Internal GND………………………….………INT. GND
● Comparator with external output
You can set the upper and lower limits, and display one of 3
results: Hi, IN or Lo. In addition to LED and buzzer results,
open collector output results are provided through the external
input/output terminals.
X: measurement value, H: Upper limit, L: Lower limit
X > H…………………………Hi
H > X > L…………………IN
L > X………………Lo
With external control
■ Save/Load function for rapid response to various
work situations
You can save and recall a maximum of 30 DMM setting
conditions for various range and comparator values.
Connector used: 57RE-40360-730B (D29) (DDK Ltd.)
Conforming connector: ADS-HC360001-010 (Honda Tsushin Kogyo Co., Ltd.),
RC30-36P (Hirose Electric Co., Ltd.) and other suitable connectors
Input type: CMOS level
Signal voltage: Hi 3.8 V to 5.0 V, Lo 0 V (shorted) to 1.2 V
Logic: Negative logic (Low active)
Output type: Open collector
Applied voltage: Max. 35 V DC, 50 mA
Logic: Negative logic (Low active)
With free running measurement
Time
MIN. TYP.
T1 Measurement trigger pulse width 500 µs …
MAX.
…
T2
Trigger delay time
See below
T3
Sampling time
using external
control
FAST
MEDIUM
See the table
SLOW at the top right of page 1.
T4 Internal operation time
… 2.0 ms …
T5
From the end of measurement
until the next trigger
500
µs
…
…
EOC Lo level time FAST … 1.7 ms …
T6 for free running MEDIUM … 50 ms …
measurement SLOW … 500 ms …
Sampling time FAST
T7 for free running MEDIUM
See the table
measurement SLOW at the top right of page 1.
■ A trigger delay designed for measurement safety
The 3237, 3238 and 3239 are equipped with a trigger delay
function that can be set to manual or automatic for the time
period between trigger input and the display of the comparator
result (see T2 in the figure above).
Manual settings: Designate periods in terms of millisecond intervals between 0.000 s and 9.999 s
Automatic settings:
FAST
MEDIUM
SLOW
DC V
3 ms
3 ms
3 ms
AC V
500 ms
800 ms
1.5 s
Ω (200Ω to 200 kΩ)
3 ms
3 ms
3 ms