10
9624 / Specifications -10
(Items in blue ■ symbolize specifications unique to Model 9624-10, and are not available in the standard Model 9624.)
-1. Function specifications
Data loading functions
Data that can be loaded : Binary data recorded using the 3196
SET files (Settings data), ITV files (TIME PLOT data),
EVT files (Event data (lists, voltage and current waveforms,
transient waveforms, numerical values), FLC files (Flicker
Maximum data capacity
Data display functions
SYSTEM display function
data (∆V10, IEC)), TRN files (Transient waveforms),
EN50160. files (EN50160 dasta),
EVENT.EN files (EN50160 Event data)
: Up to 528 MB
Screen display
: SYSTEM (settings) content screen
TIME PLOT display function
Screen display
: RMS fluctuation, voltage fluctuation, harmonic
fluctuation, inter-harmonic fluctuation
Number of display screens : Up to 4 screens
Cursor function
: A and B cursors (interval calculation function provided)
EVENT list display function
Screen display
: EVENT list content display
Display method selection : Order events occurred in, or order of priority
EVENT data display function
Display function
: Display the event data selected in the EVENT list
display screen
Screen display
: Display one of the following screens ((1) to (4))
(1)Waveform display :Select from the voltage/current waveform, 4-channel
voltage waveform, 4-channel current waveform, and
voltage/transient overvoltage waveform displays.
(2)Vector display :Select from the harmonic RMS value and phase angle
displays.
(3)DMM display :Displays power, voltage, and current values.
(4)Harmonics display :Select from the harmonics bar graph and list displays.
Cursor function
: A and B cursors (interval calculation function provided)
for the waveform display screen
Positive/ Negative/ Zero phase calculation function
:Display voltage and current of the positive phase, negative
phase, and zero phase.(In vector display screen, this is conducted
during the 3P4W wiring analysis.)
Flicker graph Display function
Screen display
: ∆V10 Flicker graph or IEC Flicker graph
Cursor function
: A and B cursors (interval calculation function provided)
EVENT voltage fluctuation graph Display function
Cursor function
: A and B cursors (interval calculation function provided)
Integral power consumption calculation function
Settings : Analysis start time/period : Set the year, month, day, hours, minutes,
and seconds. /1 to 31 days
Display method and calculation items :
Integral power consumption graph, Integral power consumption (consumption +
regeneration, and cursor measurement functions provided), Maximum integral power
consumption (final integral power consumption for the specified analysis period)
Demand calculation function
Settings :Analysis start time/period : Set the year, month, day, hours, minutes,
and seconds. /1 to 31 days
Demand interval settings : 5, 10, or 30 minutes, 1, 2, 3, 6, or 12 hours
Display method and calculation items :
Demand graph (for consumption only), Average demand (average demand value
for the specified analysis period), Maximum demand (maximum demand value for the
specified analysis period), Load ratio (average demand/maximum demand × 100 [%])
ITIC curve display function
Display function
: Plot event points on limited value curve (points indicating
swell/dip/interruption occurrence period and voltage)
Percent of nominal voltage: Maximum swell voltage or residual voltage ratio against
official voltage
Violation count display : Number of upper-limit violations, number of lower- limit
violations, and total number of events
Limit curve selection
: ITIC curve or user-defined curve (any setting)
EN50160 display function
Screen display
: Overview/Harmonic/Signaling/Measurement result sorting
Copy function
Copy content : Saves the various screens in BMP format
Print function
Print format: Prints screen images, Paper size: A4 and Letter, Print preview: Yes
CSV format conversion function
Screens that can be converted : TIME PLOT and event waveform screens
Conversion settings
: Specified interval conversion (TIME PLOT screen only)
Conversion setting selection (TIME PLOT screen only)
Report creation function
Output format : Output setting contents can be printed, or saved as a rich text file.
(1) Automatic output : RMS voltage fluctuation graph, worst case, maximum/minimum
list, total harmonic voltage distortion graph, Overview and
Signaling data of EN50160, and all event detail list.
(2) Arbitrary output : Includes, in addition to automatic output, RMS current
fluctuation graph, transient waveform, total harmonic current
distortion graph, Harmonic and result classification data of the
EN50160, and settings list.
(3) Detailed output : Voltage fluctuation, RMS fluctuation, harmonics fluctuation, and
interharmonics fluctuation.
Settings save function
Save user-defined curves, setting for sorting measurement result, report setting, etc.
Download function
Download data from the 3196 via LAN.
-2. Basic specifications
Supplied accessories : CD-R × 1
Operating environment : PC/AT-compatible devices
OS
: English or Japanese versions of the following
• Microsoft Windows 95 (9624 only, OSR2 or later versions
only supported, Internet Explorer 3 or later required)
• Microsoft Windows 98, Me, NT 4.0, 2000 or XP
Memory
: At least 128 MB
3196 Specifications
-1. Measurement and recording items
Recording item Power P&Harm ALL_D Recording item Power P&Harm ALL_D
Transient overvoltage
❍ ❍ ❍ Voltage unbalance factor ❍ ❍ ❍
Voltage swell
Voltage dip
Instantaneous interruption
Frequency
RMS voltage
RMS current
Voltage peak
❍ ❍ ❍ Current unbalance factor ❍ ❍ ❍
❍ ❍ ❍ Harmonic voltage
× ❍❍
❍ ❍ ❍ Harmonic current
× ❍❍
❍ ❍ ❍ Harmonic power
× ❍❍
❍ ❍ ❍ Harmonic voltage-current phase difference × ❍ ❍
❍ ❍ ❍ Inter-harmonic voltage
× ×❍
❍ ❍ ❍ Inter-harmonic current
× ×❍
Current peak
❍ ❍ ❍ Total harmonic voltage distortion factor ❍ ❍ ❍
Effective power
Apparent power
Reactive power
❍ ❍ ❍ Total harmonic current distortion factor ❍ ❍ ❍
❍ ❍ ❍ Total inter-harmonic voltage distortion factor × × ❍
❍ ❍ ❍ Total inter-harmonic current distortion factor × × ❍
❍ Power factor/Displacement power factor ❍ ❍ K factor
❍❍❍
Flicker (∆V10/Pst, Plt)
❍❍❍
* Select from a total of six different patterns when recording data. These consist of three available data patterns
(Power, P&Harm, or ALL DATA), combined with two patterns, AVE and ALL (maximum, minimum, and average), of
detailed data for each measurement item.
-2. Basic specifications
Power quality measurement
standards conformance : IEC61000-4-30:2002, IEEE1159,
EN50160:1999
Clock functions : Auto calendar, auto leap year, 24-hour clock
Real-time clock accuracy : Within ±0.3 s/day (when the 3196 is turned on)
Internal memory capacity for data : 13 MB (time series and event data)
Maximum recording interval : 1 month (internal memory)
Measurement time control : Manual/Specified time
Time series data settings
Recording item setting patterns : Power, P&Harm, and ALL DATA
MAX/MIN/AVE values
: AVE values, ALL values (maximum,
minimum, and average values)
Interval selections
: 1, 3, 15, or 30 seconds, 1, 5, 10, 15, or
30 minutes, 1 or 2 hours
Event settings
Event settings
: All measurement settings except
flicker and inter-harmonics
Event threshold value setting : OFF or desired numerical value
Maximum number of recording events : 100 (internal memory)
(Simultaneous events count as 1 event.)
Power supply : 12 V DC from the 9458 AC ADAPTER or 9459 BATTERY PACK
Maximum rated power : 40 VA
Continuous operating time with battery : Approximately 30 minutes
(9459 battery pack)
External dimensions : Approximately 298W (11.73") × 215H (8.46") × 67D
(2.64") mm (not including projections)
Mass : Approximately 2.25 kg (79.4 oz.) (including 9459 battery pack)