English
Language : 

DS-0027 Datasheet, PDF (2/2 Pages) All Sensors Corporation – -10 cm H2O to120 cm H2O Pressure Sensor
Pressure Sensor Characteristics Maximum Ratings
Supply Supply Voltage VS
Common-mode pressure
Lead Temperature
(soldering 2-4 sec.)
+4.5 to +5.5 Vdc
10 psig
250°C
Standard Pressure Ranges
Environmental Specifications
Temperature Ranges
Compensated
Operating
Storage
Humidity Limits
see specification
-25 to 85° C
-40 to 125° C
0 to 95% RH
(non condensing)
Part Number
120 CMH2O-D-4V
120 CMH2O-D-4V-PRIME
Operating Pressure
-10 to 120 cm H2O
-10 to 120 cm H2O
Compensated Range
5 to 50° C
-25 to 85° C
Proof Pressure
300 cmH2O
300 cmH2O
Burst Pressure
600 cmH2O
600 cmH2O
Performance Characteristics for 120 CMH2O-D-4V
Parameter, note 1
Operating Range, pressure
Output Voltage, @120 cmH2O
Output Voltage @ zero pressure
Output Voltage @-10 cmH2O
Offset Temperature Shift (5°C-50°C), note 2
Linearity, hysteresis error, note 4
Span Shift (5°C-50°C), note 2
Minimum
-10
4.40
0.30
0.23
--
--
--
Nominal
--
4.5
0.35
0.28
--
0.05
--
Maximum
120
4.6
0.40
0.33
±1.0
±0.25
±1.0
Units
cmH2O
volt
volt
volt
%span
%fs
%span
Performance Characteristics for 120 CMH2O-D-4V-PRIME
Parameter, note 1
Operating Range, pressure
Output Voltage, @120 cmH2O
Output Voltage @ zero pressure
Output Voltage @-10 cmH2O
Offset Temperature Shift (-25 to 85°C), note 2
Linearity, hysteresis error, note 4
Span Shift (-25 to 85°C), note 2
Minimum
-10
4.40
0.30
0.23
--
--
--
Nominal
--
4.5
0.35
0.28
--
0.05
--
Maximum
120
4.6
0.40
0.33
±1.0
±0.25
±1.0
Units
cmH2O
volt
volt
volt
%span
%fs
%span
Specification Notes
note 1: All parameters are measured at 5.0 volt excitation, for the nominal full scale pressure and room temperature unless
otherwise specified. Pressure measurements are with positive pressure applied to PORT B.
note 2: Shift is relative to 25°C.
note 3: Shift is within the first hour of excitation applied to the device.
note 4: Measured at one-half full scale rated pressure using best straight line curve fit.
note 5: The voltage added to the offset voltage at full scale pressure. Nominally the output voltage range is 0.35V to 4.5V
volts for minus to plus full scale pressure.
All Sensors reserves the right to make changes to any products herein. All Sensors does not assume any liability arising out of the application or use of
any product or circuit described herein, neither does it convey any license under its patent rights nor the rights of others.
Amplified Pressure Sensors