English
Language : 

NLP110 Datasheet, PDF (1/3 Pages) Artesyn Technologies – LOW TO MEDIUM POWER AC/DC POWER SUPPLIES
NLP110 Medical Series
Single and triple output
LOW TO MEDIUM POWER AC/DC POWER SUPPLIES 80-110W AC/DC Universal Input Switch Mode Power Supplies 1
• Provides low voltage outputs
• Compliance to EN61000-3-2 (Power Factor = 0.98)
• Smallest industry standard package 6.5 x 3.0 x 1.26 inches
• 90VAC to 264VAC universal input range
• UL, cUL, and VDE safety approvals and CE-Marked to MDD
• Overvoltage and short circuit protection
• VDE0871-A, EN55022-A, EN55011-A conducted noise
• VDE0871-A, EN55022-A, EN55011-A radiated noise
• Meets all applicable and relevant immunity standards
EN61000-4-2,4,5 and 6
The NLP110 Medical is a series of 110W AC/DC open-frame power supplies in a
3” x 6.5” x 1.26” package. The NLP110 Medical is power factor corrected to meet EN61000-3-2
and is available in single and multiple output models. The multiple output models have a capability
that is unique to the power supply market - they can deliver equal amounts of current on multiple
low voltage outputs to support designs that will span the industry transition from 5V logic to 3.3V
logic. The NLP110 Medical is practically the same size as a 65 Watt 3” x 5” product, e.g. NLP65
Medical - the one difference is that the footprint is a mere 1.5 inches longer. The NLP110 Medical
series, with full medical safety approval per UL2601-1, C22.2 No. 601.1-M90, VDE0750,
IEC60601-1 and EN60601-1 and CE marked, greatly accelerating design-in time and reducing
system compliance costs.
All specifications are typical at nominal input, full load at 25°C unless otherwise stated
2 YEAR WARRANTY
SPECIFICATIONS
OUTPUT SPECIFICATIONS
Total regulation
(Line and load)
Main output
Auxiliary outputs
±2.0%
±5.0%
Rise time
At turn-on
1.0s, max.
Transient response
Main output
75% to 100%
step at 0.1A/µs
5.0% or 250mV
max. dev., 1ms max.
recovery to 1%
Temperature coefficient
±0.02%/°C
Overvoltage protection Main outputs
125%, ±10%
Short circuit protection Cyclic operation
Continuous
Minimum output current Single and multiple
See table
INPUT SPECIFICATIONS
Input voltage range
(See Note 6)
Universal input
90 to 264VAC
Input frequency range
47Hz to 63Hz
Input surge current
(cold start)
120VAC
230VAC
18A max.
35A max.
Safety ground
leakage current
230VAC, 50Hz
<100µA
Input current
120VAC @ 80W
120VAC @ 110W
230VAC @ 80W
230VAC @ 110W
0.9A rms
1.3A rms
0.48A rms
0.7A rms
Input fuse
UL/IEC127
F3.15A H, 250VAC
EMC CHARACTERISTICS (12)
Conducted emissions
Radiated emissions
Harmonic current
emission correction
ESD air
ESD contact
EN55022, FCC part 15
EN55022, FCC part 15
EN61000-3-2
EN61000-4-2
EN61000-4-2
Level A
Level A
Compliant
Level 3
Level 3
EMC CHARACTERISTICS (continued)
Surge
Fast transients
Radiated immunity
Conducted immunity
EN61000-4-5
EN61000-4-4
EN61000-4-3
EN61000-4-6
Level 3
Level 3
Level 3
Level 3
GENERAL SPECIFICATIONS
Hold-up time
120VAC, 60Hz
16.7ms @ 110W
Efficiency
120VAC, 80W (-9605)
75% min.
Isolation voltage
Input/output
Input/chassis
4000VAC
1500VAC
Approvals and
standards
EN60601-1, IEC60601-1,VDE0750,
UL2601-1, C22.2 No. 601.1-M90
Weight
283g (10oz)
MTBF (@ 25°C)
MIL-HDBK-217F 220,000 hours min.
ENVIRONMENTAL SPECIFICATIONS (9)
Thermal performance
Operating ambient,
0°C to +50°C
(See derating curve)
Non-operating
-40°C to +85°C
50°C to 70°C, ambient,
Derate to
convection cooled
50% load
0°C to 50°C, ambient,
80W
convection cooled
0°C to 50°C ambient,
110W
150LFM forced air
Peak (0°C to +50°C, 60s) (See Note 3)
Relative humidity
Non-condensing
5% to 95% RH
Altitude
Operating
Non-operating
10,000 feet max.
30,000 feet max.
Vibration (See Note 8) 5Hz to 500Hz
2.4G rms peak
Shock
per MIL-STD-810E
516.4 Part IV
File Name: NLP110M.PDF Rev: 13 Dec 2002