English
Language : 

MTC_SERIES Datasheet, PDF (2/3 Pages) Arch Electronics Corp. – Fully Encapsulated Plastic Case
MTC SERIES
VER : B_5 update : 2011.01.07
AC-DC Medical
15 Watts
Input
Output
Protection
Isolation
Environment
Physical
Safety & EMC
Voltage
Frequency (Hz)
Current (Full load)
Inrush current (<2ms)
Leakage Current
External fuse (recommend)
Ripple & Noise
Hold-up time
Switching frequency
Over current protection
Over voltage protection
Short circuit protection
OTP
Input-Output (V.AC)
Operating temperature
Storage temperature
Temperature coefficient
Humidity
MTBF
Dimension (L x W x H)
Case Material
Weight
Cooling method
Safety Approvals
approval
EMI / PFI conducted
EMC Standards
ESD Susceptibility
Radiated Susceptibility
EFT / Burst
Surge
Conducted Susceptibility
90-264 VAC or 100-375 VDC
47-440 Hz
220 mA max. (115 VAC) / 118 mA max. (230 VAC)
10 A max. (115 VAC) / 20 A max. (230 VAC)
0.1 mA max. (115 VAC) / 0.2 mA max. (230 VAC)
2.0 A slow blow type
1% of Vout
20 ms min.
132 kHz
Above 105% rated output power
Zener diode clamp
Hiccup mode, indefinite (automatic recovery)
100°C
4000V
-25°C…+70°C (Case Temperature max. +95°C)
-40°C…+100°C
0.01%/°C
95% RH
200,000 h~400,000 h @ 25°C
2.52 x 1.79 x 0.92 Inches ( 64.0 x 45.6 x 23.5 mm ) Tolerance ±0.5 mm
Plastic resin + Fiberglass (flammability to UL 94V-0)
120 g
Free air convection
UL60601-1 / IEC60601-1
cUL / UL / CB
EN 55011 class B
EN60601-1-2
EN60601-1-2
EN60601-1-2
EN60601-1-2
EN60601-1-2
EN60601-1-2
1. All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
2. Ripple & Noise are measured at 20MHz of bandwidth with 0.1μF & 47μF parallel capacitor.