English
Language : 

MQF240E Datasheet, PDF (2/3 Pages) Arch Electronics Corp. – Enclosed Medical Switching Power Supply
MQF240E SERIES
VER : PA_3 update : 2016.07.21
AC-DC Medical Switching Power Supply
240 Watts
ELECTRICAL SPECIFICATIONS
All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated.
Model No.
Safety
Approval
Conducted and radiated EMI
MQF240E-12S
MQF240E-24S
MQF240E-48S
cUL / UL Standard:
UL 60950-1, CAN/CSA C22.2 No. 60950-1-07
ANSI/AAMI ES60601-1 (2005 + C1:09 + A2:10), CAN/CSA-C22.2 No. 60601-1
(2008), 2 x MOPP (Input to Output)
EN55011 class B (Pending)
ESD
EN61000-4-2 air ± 8kV , Contact ± 4Kv (Pending)
Radiated Immunity
EN61000-4-3 10V/m (Pending)
Fast Transient
EN61000-4-4 ± 2kV (Pending)
EMC
Surge
EN61000-4-5 ±1kV (Pending)
Conducted Immunity
EN61000-4-6 10Vrms (Pending)
PFMF
EN61000-4-8 30A/m (Pending)
Dips
EN61000-4-11 30% 10ms (Pending)
Interruption
EN61000-4-11 >95% 5000ms (Pending)
NOTE
1. Ripple & Noise are measured at 20MHz of bandwidth with 0.1uF & 47uF parallel capacitor.
2. Hold-up Time measured at 90% Vout.
3. Main Vout must be >50% Load, 12V (Aux) / 0.5A.
4. Please check the derating curve for more details.
5. Strongly recommend to conduct this test with DC Voltage. If customer wishes to test with AC Voltage,
please disconnect all Y-Capacitors within Arch power supply.
DERATING
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
-2-