English
Language : 

MQF500E Datasheet, PDF (1/4 Pages) Arch Electronics Corp. – AC-DC Medical Switching Power Supply
MQF500E SERIES
VER : A_0 update : 2016.07.14
AC-DC Medical Switching Power Supply
500 Watts
KEY FEATURES
 Enclosed Medical Switching Power Supply
 Remote ON/OFF Function
 Built-in 12V/0.3A Auxiliary Output
 Standby 5V@1A
 High Efficiency up to 92%
 With P.F.C. Function >0.95
 Ultra Compact Size: 5.5 x 3.32 x 2.44 Inches
 3-Year Product Warranty
ELECTRICAL SPECIFICATIONS
(Pending)
All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated.
Model No.
Max Output Wattage (W)
Voltage
MQF500E-12S
MQF500E-24S
500 W
90-264 VAC or 127-370 VDC
MQF500E-48S
Frequency (Hz)
47-63 Hz
Input
Current (Full load)
Inrush Current (<2ms) (Clod Start)
<6.3 A max. (115 VAC) / <3.15 A max. (230 VAC)
< 40 A max. (115 VAC) / < 80 A max. (230 VAC)
Leakage Current
< 0.1 mA max. (Input-Output)
Power Factor (at 230 VAC)
PF>0.95 at Full Load
Voltage (V.DC.)
12V
24V
48V
Voltage Accuracy
±2%
Voltage Adj. Range (V.DC)
11.52~12.48
23.04~24.96
46.08~49.44
Current (A) max
41.5
20.8
10.41
Output
Line Regulation (115-264 VAC)
Load Regulation (10-100%) (typ.)
±0.5%
±1%
Minimum Load
3%
Maximum Capacitive Load
10,000µF
5,000µF
2,500µF
Ripple & Noise (typ.)
160mV
240mV
480mV
Efficiency (at 230 VAC)
89%
91%
92%
Hold-up Time (at 115 VAC)
8 ms min.
Over Power Protection
Auto recovery
Protection
Over Voltage Protection
Overt Temperature Protection
Auto recovery
Auto recovery
Short Circuit Protection
Auto recovery
Isolation
Input-Output (V.AC)
Input-FG (V.AC)
4000VAC or 5656VDC
2000V
Output-FG .(V.AC)
1500V
Operating Temperature
-30°C…+70°C (with derating)
Storage Temperature
-35°C…+85°C
Temperature Coefficient
Environment
±0.03%/°C ( 0~50°C )
±0.06%/°C ( -30~0°C )
Humidity
95% RH
MTBF
>160,000 h @ 25°C (MIL-HDBK-217F)
Vibration
10~500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes.
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
-1-