DIMENSIONS AND WEIGHT
Dimensions
3010/11 Single width 3U PXI module 3030A Double width 3U PXI
module
Weight
3010 375 g (0.8 lbs)
3011 390 g (0.86 lbs)
3030A 750 g (1.7 lbs)
FREQUENCY REFERENCE IN
Mode
External, Free Running (3030A)
External Source
SMA
Frequency
10 MHz
Level
0.4 V to 4 V pk-pk into 50 ohms or looped through
GENERAL
The following general specifications are common to the 3010, 3011
and 3030A
Standard Warranty
24 months
Calibration Interval
Recommended 24 months
Electromagnetic Compatibility
EN 61326-1:1997, Emissions Class A, Immunity Table 1 -
Performance Criteria B
Safety
EN 61010-1:2001 Safety requirements for electrical equipment for
measurement, control and laboratory use-Part 1, General requirements
Driver Software
VXIpnp compliant software driver
LOCAL OSCILLATOR
Frequency range
1500 MHz to 3000 MHz
Level
3010 output: -4 dBm to +3 dBm 3030A input: Nominally 0 dBm
RATED RANGE OF USE
Operating Temperature
0 to 50°C, meets IEC-60068-2-1 and 60068-2-2
Operating Humidity
10 to 90% non-condensing, meets IEC-60068-2-56
CONDITIONS OF STORAGE AND TRANSPORT
Storage Temperature
-20 to +70°C, meets IEC-60068-2-1 and 60068-2-2
Storage Humidity
5 to 93% non-condensing, tested to IEC-60068-2-56
Shock
30 g peak, half sine, 9 ms pulse. Tested in accordance with IEC-
60068-2-27
Random vibration 5 Hz to 500 Hz, 2.46 g rms non-operating. Tested
in accordance with IEC-60068-2-64
COMPLIANCE
PXI Specification, Revision 2.1
VXIplug&play Specifications (VPP-2, VPP-3.x, VPP-4.x and VPP-7)
3010/3011 SPECIFIC SPECIFICATIONS
Specifications are common to the 3010 and 3011 unless otherwise
stated.
LO OUTPUT
Frequency Range
1.5 GHz to 3.0 GHz
Resolution
1 Hz
Accuracy
As frequency standard
Output Power
Fixed level in the range -4 dBm to +3 dBm
Output Impedance
50 Ω nominal
VSWR
<2:1
FREQUENCY REFERENCE OUT (3011 ONLY)
Level
2 V pk-pk nominal square wave into 50 ohms
Frequency
10 MHz
Aging Rate
1 in 109 per day, 1 in 107 per year
Temperature Stability (0 to 50°C)
Typically better than ±1 x 10-8
Warm-Up Time
<5 Minutes