ADAM-5018 Datasheet, PDF (1/1 Pages) Advantech Co., Ltd. – 8-ch Ultra High Speed Analog Input | |||
|
ADAM-5017UH
ADAM-5018
ADAM-5018P
8-ch Ultra High Speed Analog Input
Module
7-ch Thermocouple Input Module
7-ch Thermocouple Input Module
with Independent Input Range
ADAM-5017UH
RoHS
COMPLIANT
2002/95/EC
ADAM-5018
RoHS
COMPLIANT
2002/95/EC
ADAM-5018P
RoHS
COMPLIANT
2002/95/EC
Specifications
General
ï§ Certifications
ï§ Connectors
ï§ Power Consumption
CE
1 x Plug-in screw terminal
(# 14 ~ 22 AWG)
2.2 W (max.)
Analog Input
ï§ Accuracy
±0.1% or better
ï§ Bandwidth
200 kHz
ï§ Channels
8 differential
ï§ Differential Non-linear ±1 LSB
ï§ Input Impedance
Voltage: 2 Mâ¦
Current: 120â¦
(Build-in 12â¦. register for
Current Input)
ï§ Integral Non-linear ±1 LSB
ï§ Input Type
mV, V, mA
ï§ Input Range
±10 V, +0 ~ 10 V,
0 ~ 20 mA, +4 ~ 20 mA
ï§ Low pass filter
Configured by User
ï§ Resolution
12-bit
ï§ Sampling Rate
Depends on base unit
ADAM-5000/485 & 5000E:
100 Samples/sec max (Total): one ADAM-5017UH
installed
ADAM-5000/TCP:
100 Samples/sec max (Total) :one ADAM-5017UH
installed
ADAM-5510 :
200K Samples/sec max (Single Channel): one
ADAM-5017UH installed
ADAM-5550 :
1K Samples/sec per channel: one ADAM-5017UH
installed
*Depending on the performance of client server or
controller
ï§ Signal Input
200 kHz for both voltage
Bandwidth
and current inputs
Protection
ï§ Isolation Voltage
Note:
3,000 VDC
1) The voltage difference between any two pins must not
exceed 15 V
2) Distinct range settings allowed on each channel
Note: Not support ADAM-5510KW series
Specifications
General
ï§ Certifications
CE, FM
ï§ Connectors
1 x Plug-in screw terminal
(# 14 ~ 22 AWG)
ï§ Power Consumption 0.63 W (max.)
Thermocouple Input
ï§ Accuracy
ï§ Bandwidth
ï§ Channels
ï§ CMR @ 50/60 Hz
ï§ Input Impedance
ï§ lnput Range
±0.1% or better
13.1 Hz @ 50 Hz
15.72 Hz @ 60 Hz
7 differential
92 dB min
2 Mâ¦
±15 mV, ±50 mV, ±100 mV,
±500 mV,
±1 V, ±2.5 V, ±20 mA
ï§ lnput Type
mV, V, mA,thermocouple
ï§ Resolution
16-bit
ï§ Sampling Rate
10 samples/sec. (total)
ï§ Span Drift
±25 PPM/ ° C
ï§ Zero Drift
±6 μV/ ° C
ï§ T/C Type and Temperature Range
J
0°
~
760° C
K
0°
~
1,370° C
T
-100°
~
400° C
E
0°
~
1,000° C
R
500°
~
1,750° C
S
500°
~
1,750° C
B
500°
~
1,800° C
Protection
ï§ Fault and
Overvoltage
Protection
ï§ Isolation Voltage
Withstands overvoltage up
to ±35 V
3,000 VDC
Ordering Information
ï§ ADAM-5018
7-ch Thermocouple Input
Module
Ordering Information
ï§ ADAM-5017UH
8-ch Ultra High Speed
Analog Input Module
Specifications
General
ï§ Certifications
CE
ï§ Connectors
1 x Plug-in screw terminal
(# 14 ~ 22 AWG)
ï§ Power Consumption 0.63 W (max.)
Thermocouple Input
ï§ Accuracy Bandwidth ±0.1% or better
13.1 Hz @ 50 Hz
15.72 Hz @ 60 Hz
ï§ Channels
7 differential with
Independent Input Range
ï§ CMR @ 50/60 Hz 92 dB min
ï§ Input Impedance 20 Mâ¦. (Build-in 120 â¦
Register for Current Input)
ï§ lnput Range and Types
Thermocouple
J
0
~
760° C
K
0
~
1,370° C
T
-100
~
400° C
E
0
~
1,000° C
R
500
~
1,750° C
S
500
~
1,750° C
B
500
~
1,800° C
ï§ Current mode
±20 mA, 4 ~ 20 mA
ï§ Voltage
±15 mV (0.2%), ±50 mV,
±100 mV, ±500 mV
±1 V, ±2.5 V
ï§ Resolution
16-bit
ï§ Sampling Rate
10 samples/sec. (total)
ï§ Span Drift
±25 PPM/°C
ï§ Zero Drift
±6 μV/°C
ï§ High Common Mode 200 VDC
Protection
ï§ Fault and
Withstands overvoltage up
Overvoltage
to ±35 V
Protection
ï§ Isolation Voltage
ï§ Filter function
3,000 VDC
Yes
ï§ Built-in TVS/ESD Protection
Note: Not support ADAM-5510KW series
Ordering Information
ï§ ADAM-5018P
7-ch Thermo. Input Module
w/ Ind. Input Range
PC-based Controllers & I/O Modules
|